Products  \  Spectrophotometer  \  Details
Q6 Dual Beam UV Visible Spectrophotometer
    Publish time 2022-12-08 12:59    




[Features] 




Advanced Technics Design


Precision injection molding case promises robust structure and longer lifetime;

Modern & smart design attracts more attention.

 

Environmental Care


Compact structure, 30% installation space saved;

Strong built-in cooling fan, 10% energy saved.

 

High Stability & Reliability


Imported high-quality deuterium lamp & tungsten lamp.

A mercury lamp equipped, one-key calibration for wavelength accuracy.

Unique C-T optical structure, ultra-low stray light;

Premium lens and lens coating, high repeatability;

Real-time automatic calibration of dark current;

Equipped with computer simulation optimized optical base;

Dual beam optical path system, lower influence on test results;

Multiple curve fitting makes curve regression more accurate.

 

Sensitive Detection


High-quality photomultiplier, fast response, high sensitivity.

Especially suitable for weak radiation energy detection.

 

Adjustable Bandwidth


Continuous adjustable spectral bandwidth from 0.1nm to 5nm;

0.1nm variable interval, suitable for various requirements;

Especially for samples with sharp absorption peaks, like penicillin sodium, penicillin potassium, etc. 

 

Multi-application  


Various optional accessories, like automatic 8-posotion cell holder, multi-function automatic sampler, thermostatic cell holder, integrating sphere, mirror reflection accessory, variable optical path cell holder and adjustable solid cell holder, etc.

Extends the application range, combining the requirements of common samples and special ones.

 

Efficient and Convenient


Higher efficiency with optional auto 8-position cell holder and high throughput 100-position autosampler.

User-friendly software interface.

Comply with GLP/GMP standards, user management, logging and data tracking are available.

Independent modular design, lower maintenance cost. 

Socket type of lamps, easy replacement and no need to adjust optical path.  

Large sample chamber, enough room for operating and replacing sample.

 

Professional Operating Software


Follow GLP/GMP standards, user management, logging, data tracking, etc.are all available.




[Technical Specification]





 

Model

Q-6

 Optical System

Double beam

Detector

PMT

Light Sources

Deuterium Lamp, Tungsten Lamp, Mercury Lamp 

 Control Mode

PC software controlled

Wavelength Range

190nm-900nm

Spectral Bandwidth

0.1nm-5nm(0.1nm interval continuously adjustable)

Wavelength Accuracy

±0.3nm

Photometric Accuracy

±0.3 %

Stray Light

≤0.005%(@220nm&360nm)

Baseline Flatness

±0.0005 Abs(190nm-850nm)

Stability

≤0.0001 Abs/h

Power Supply

AC 220 V/ AC 110 V, 50/60Hz, 500

Dimensions(L*W*H)

500*550*260 mm